Tektronix DMM4050


0 VNĐ

Hotline đặt hàng08 3601 6797



Đồng Hồ Đo Vạn Năng Tektronix DMM4050

Đồng hồ đo vạn năng, đồng hồ vạn năng Tektronix DMM4050 phù hợp với nhiều loại đo trong lĩnh vực điện tử viễn thông.

Mô tả Đồng hồ VOM Tektronix DMM4050

Voltage

Characteristic
Description
100 V Setting
90 V to 110 V
120 V Setting
108 V to 132 V
220 V Setting
198 V to 242 V
240 V Setting
216 V to 264 V
Frequency
47 Hz to 440 Hz. Automatically sensed at power-on
Power Consumption
28 VA peak (12 Watt average)
Dimensions
Dimension
mm
in.
Height
88
3.46
Width
217
8.56
Depth
297
11.7
Weight
kg
lb.
Net
3.6
8.0
Shipping
5.0
11.0
Display
Vacuum Fluorescent Display, dot matrix

Environment

Characteristic
Description
Temperature
   Operating
0 °C to 55 °C
   Storage
–40 °C to 70 °C
   Warm Up
1 hour to full uncertainty specifications
Relative Humidity (noncondensing)
   Operating
<90% (0 °C to 28 °C)
<80% (28 °C to 40 °C)
<50% (40 °C to 55 °C)
   Storage
<95% (–40 °C to 70 °C)
Altitude
   Operating
2,000 meters
   Storage
12,000 meters
Vibration
Complies with Mil-T-28800F Type III, Class 5 (Sine only)
Safety
Designed to comply with IEC 61010-1:2000-1, UL 61010-1A1, CAN/CSA-C22.2 No. 61010.1, CAT I 1000V / CAT II 600 V
EMC
Designed to comply with IEC 61326-1:2000-11 (EMC) when used with shielded communications cables. This meter has shown susceptibility to radiated frequencies greater than 1 V/m from 250 to 450 MHz

Triggering

Characteristic
Description
Samples per Trigger
1 to 50,000
Trigger Delay
0 s to 3600 s; in 10 μs increments
External Trigger Delay
<1 ms
External Trigger Jitter
<500 μs
Trigger Input
TTL Levels
Trigger Output
5 V max (open collector)
Memory
Model
Description
DMM4050/4040
10,000 measurements, internal, and up to 2 Gigabyte capacity with USBmemory module (available separately) through front-panel USB port

Math Functions

Zero, dBm, dB, MX+B, Offset, DCV ratio and TrendPlot, Histogram, Statistics (min/max/average/standard deviation), and Limit Test

Dual Function Compatibility Matrix

Primary Function
Secondary Function
VDC
VAC
IDC
IAC
FREQ
OHMS
DIODE
Period
RTD
VDC

X







VAC
X



X




IDC



X





IAC


X

X




FREQ







X

OHMS








X
DIODE









Continuity




X




Electrical
Characteristic
Description
Input Protection
1000 V all ranges
Overrange
20% on all ranges except 1000 VDC, 1000 VAC, Diode, and 10 A ranges
Remote Interfaces
RS-232C, DTE 9-pin, 1200 to 230400 baud (RS-232C to USB cable available to connect the meter to a PC USB port) IEEE 488.2.
LAN and “Ethernet 10/100BASE-T with DHCP (for IP address) option”.

Warranty

Three years

ELECTRICAL SPECIFICATIONS
DC VOLTAGE SPECIFICATIONS

Accuracy specifications are valid for 6½ digit resolution mode after at least a 1 hour warm-up with Auto Zero enabled.
24-hour specifications are relative to calibration standards and assume a controlled electromagnetic environment per EN 61326-1:2000-11.
Characteristic
Description
Maximum Input
1000 V on any range
Common Mode Rejection
140 dB at 50 or 60 Hz ±0.1% (1 kΩ unbalance)
Normal Mode Rejection
60 dB for NPLC of 1 or greater with analog filter off and power line frequency ±0.1%
100 dB for NPLC of 1 or greater with analog filter on and power line frequency ±0.1%
Measurement Method
Multiramp A/D
A/D Linearity
0.0002% of measurement + 0.0001% of range
Input Bias Current
<30 pA at 25 °C
Auto-zero Off Operation
Following instrument warm-up at calibration temperature ±1 °C and less than 10 minutes, add error: 0.0002% range additional error +5 μV
Analog Filter
When using the analog filter, specifications are relative to within one hour of using the ZERO function for that range and NPLC setting
DC Ratio
Accuracy is ±(Input accuracy + Reference accuracy), where Input accuracy = DC Voltage accuracy for the HI to LO Input (in ppm of the Input voltage), and Reference accuracy = DC Voltage accuracy for the HI to LO (Sense) Reference (in ppm of the Reference voltage)
Settling Considerations
Measurement settling times are affected by source impedance, cable dielectric characteristics, and input signal changes
Input Characteristics
Range
Resolution
Resolution
Input Impedance
4½ Digits
5½ Digits
6½ Digits
100 mV
100.0000 mV
10 μV
1 μV
100 nV
10 MΩ or >10 GΩ 2
1 V
1.000000 V
100 μV
10 μV
1 μV
10 MΩ or >10 GΩ 2
10 V
10.00000 V
1 mV
100 μV
10 μV
10 MΩ or >10 GΩ 2
100 V
100.0000 V
10 mV
1 mV
100 μV
10 MΩ ±1%
1000 V
1,000.000 V
100 mV
10 mV
1 mV
10 MΩ ±1%
2 Inputs beyond ±14 V are clamped through 200 kΩ typical. 10 MΩ is default input impedance.
DMM4050 Accuracy
Range
24 Hour
(23 °C ±1 °C)
90 Days
(23 °C ±5 °C)
1 Year
(23 °C ±5 °C)
Temperature Coefficient/°C
Outside 18 to 28 °C
Accuracy is given as ±(% measurement + % of range)
100 mV
0.0025 + 0.003
0.0025 + 0.0035
0.0037 + 0.0035
0.0005 + 0.0005
1 V
0.0018 + 0.0006
0.0018 + 0.0007
0.0025 + 0.0007
0.0005 + 0.0001
10 V
0.0013 + 0.0004
0.0018 + 0.0005
0.0024 + 0.0005
0.0005 + 0.0001
100 V
0.0018 + 0.0006
0.0027 + 0.0006
0.0038 + 0.0006
0.0005 + 0.0001
1000 V
0.0018 + 0.0006
0.0031 + 0.001
0.0041 + 0.001
0.0005 + 0.0001
DMM4040 Accuracy
Range
24 Hour
(23 °C ±1 °C)
90 Days
(23 °C ±5 °C)
1 Year
(23 °C ±5 °C)
Temperature Coefficient/°C
Outside 18 to 28 °C
Accuracy is given as ±(% measurement + % of range)
100 mV
0.003 + 0.003
0.004 + 0.0035
0.005 + 0.0035
0.0005 + 0.0005
1 V
0.002 + 0.0006
0.003 + 0.0007
0.004 + 0.0007
0.0005 + 0.0001
10 V
0.0015 + 0.0004
0.002 + 0.0005
0.0035 + 0.0005
0.0005 + 0.0001
100 V
0.002 + 0.0006
0.0035 + 0.0006
0.0045 + 0.0006
0.0005 + 0.0001
1000 V
0.002 + 0.0006
0.0035 + 0.0010
0.0045 + 0.0010
0.0005 + 0.0001
Additional Errors
Digits
NPLC
Additional NPLC Noise Error
100
0% of range
10
0% of range
1
0.001% of range
0.2
0.0025% of range ±12 μV
0.02
0.017% of range ±17 μV

AC VOLTAGE SPECIFICATIONS

AC Voltage specifications are for AC sinewave signals >5% of range. For inputs from 1% to 5% of range and <50 kHz, add an additional error of 0.1% of range, and for 50 kHz to 100 kHz, add 0.13% of range.
Characteristic
Description
Maximum Input
1000 VRMS or 1414 V peak or 8 × 107 Volts-Hertz product (whichever is less) for any range
Measurement Method
AC-coupled true RMS. Measures the AC component of input with up to 1000 VDC bias on any range
AC Filter Bandwidth
   Slow
3 Hz – 300 kHz
   Medium
20 Hz – 300 kHz
   Fast
200 Hz – 300 kHz
Common Mode Rejection
70 dB at 50 Hz or 60 Hz ±0.1% (1 kΩ unbalance)
Maximum Crest Factor
5:1 at Full Scale
Additional Crest Factor Errors (<100 Hz)
Crest factor 1-2, 0.05% of full scale
Crest factor 2-3, 0.2% of full scale
Crest factor 3-4, 0.4% of full scale
Crest factor 4-5, 0.5% of full scale
Only applies for non-sinusoid signals

Input Characteristics
Range
Resolution
Resolution
Input Impedance
4½ Digits
5½ Digits
6½ Digits
100 mV
100.0000 mV
10 μV
1 μV
100 nV
1 MΩ ±2% shunted by <100 pf
1 V
1.000000 V
100 μV
10 μV
1 μV
10 V
10.00000 V
1 mV
100 μV
10 μV
100 V
100.0000 V
10 mV
1 mV
100 μV
1000 V
1,000.000 V
100 mV
10 mV
1 mV
DMM4050/4040 Accuracy
Range
Frequency
24 Hour
(23 °C ±1 °C)
90 Days
(23 °C ±5 °C)
1 Year
(23 °C ±5 °C)
Temperature Coefficient/°C
Outside 18 to 28 °C
Accuracy is given as ±(% measurement + % of range)
100 mV
3 – 5 Hz
1.0 + 0.03
1.0 + 0.04
1.0 + 0.04
0.1 + 0.004
5 – 10 Hz
0.35 + 0.03
0.35 + 0.04
0.35 + 0.04
0.035 + 0.004
10 Hz – 20 kHz
0.04 + 0.03
0.05 + 0.04
0.06 + 0.04
0.005 + 0.004
20 – 50 kHz
0.1 + 0.05
0.11 + 0.05
0.12 + 0.05
0.011 + 0.005
50 – 100 kHz
0.55 + 0.08
0.6 + 0.08
0.6 + 0.08
0.06 + 0.008
100 – 300 kHz 3
4.0 + 0.50
4.0 + 0.50
4.0 + 0.50
0.20 + 0.02
1 V
3 – 5 Hz
1.0 + 0.02
1.0 + 0.03
1.0 + 0.03
0.1 + 0.003
5 – 10 Hz
0.35 + 0.02
0.35 + 0.03
0.35 + 0.03
0.035 + 0.003
10 Hz – 20 kHz
0.04 + 0.02
0.05 + 0.03
0.06 + 0.03
0.005 + 0.003
20 – 50 kHz
0.1 + 0.04
0.11 + 0.05
0.12 + 0.05
0.011 + 0.005
50 – 100 kHz
0.55 + 0.08
0.6 + 0.08
0.6 + 0.08
0.06 + 0.008
100 – 300 kHz 3
4.0 + 0.50
4.0 + 0.50
4.0 + 0.50
0.2 + 0.02
10 V
3 – 5 Hz
1.0 + 0.02
1.0 + 0.03
1.0 + 0.03
0.1 + 0.003
5 – 10 Hz
0.35 + 0.02
0.35 + 0.03
0.35 + 0.03
0.035 + 0.003
10 Hz – 20 kHz
0.04 + 0.02
0.05 + 0.03
0.06 + 0.03
0.005 + 0.003
20 – 50 kHz
0.1 + 0.04
0.11 + 0.05
0.12 + 0.05
0.011 + 0.005
50 – 100 kHz
0.55 + 0.08
0.6 + 0.08
0.6 + 0.08
0.06 + 0.008
100 – 300 kHz 3
4.0 + 0.50
4.0 + 0.50
4.0 + 0.50
0.2 + 0.02
100 V
3 – 5 Hz
1.0 + 0.02
1.0 + 0.03
1.0 + 0.03
0.1 + 0.003
5 – 10 Hz
0.35 + 0.02
0.35 + 0.03
0.35 + 0.03
0.035 + 0.003
10 Hz – 20 kHz
0.04 + 0.02
0.05 + 0.03
0.06 + 0.03
0.005 + 0.003
20 – 50 kHz
0.1 + 0.04
0.11 + 0.05
0.12 + 0.05
0.011 + 0.005
50 – 100 kHz
0.55 + 0.08
0.6 + 0.08
0.6 + 0.08
0.06 + 0.008
100 – 300 kHz 3
4.0 + 0.50
4.0 + 0.50
4.0 + 0.50
0.2 + 0.02
1000 V
3 – 5 Hz
1.0 + 0.015
1.0 + 0.0225
1.0 + 0.0225
0.1 + 0.00225
5 – 10 Hz
0.35 + 0.015
0.35 + 0.0225
0.35 + 0.0225
0.035 + 0.00225
10 Hz – 20 kHz
0.04 + 0.015
0.05 + 0.0225
0.06 + 0.0225
0.005 + 0.00225
20 – 50 kHz
0.1 + 0.03
0.11 + 0.0375
0.12 + 0.0375
0.011 + 0.00375
50 – 100 kHz 4
0.55 + 0.06
0.6 + 0.06
0.6 + 0.06
0.06 + 0.006
100 – 300 kHz 3, 4
4.0 + 0.375
4.0 + 0.375
4.0 + 0.375
0.2 + 0.015
3 Typically 30% reading error at 1 MHz.
4 1000 V range is limited to 8 × 107 Volt-Hertz.
Additional Low Frequency Errors
Frequency
AC Filter
3 Hz (Slow)
20 Hz (Medium)
200 Hz (Fast)
Error is stated as % of reading.
10 – 20 Hz
0
0.25
20 – 40 Hz
0
0.02
40 – 100 Hz
0
0.01
0.55
100 – 200 Hz
0
0
0.2
200 Hz – 1 kHz
0
0
0.02
>1 kHz
0
0
0

RESISTANCE

Specifications are for 4-wire resistance function, 2 × 4-wire resistance, or 2-wire resistance with zero. If zero is not used, add 0.2 Ω for 2-wire resistance plus lead resistance, and add 20 mΩ for 2 × 4-wire resistance function.
Characteristic
Description
Measurement Method
Current source referenced to LO input
Max Lead Resistance (4-wire ohms)
10% of range per lead for 10 Ω, 100 Ω, 1 kΩ ranges. 1 kΩ per lead on all other ranges
Input Protection
1000 V on all ranges
Common Mode Rejection
140 dB at 50 or 60 Hz ±0.1% (1 kΩ unbalance)
Normal Mode Rejection
60 dB for NPLC of 1 or greater with analog filter off and power line frequency ±0.1%
100 dB for NPLC of 1 or greater with analog filter on and power line frequency ±0.1%
Analog Filter
When using the analog filter, specifications are relative to within one hour of using the ZERO function for that range and NPLC setting

Input Characteristics
Range
Resolution
Resolution
Source Current
4½ Digits
5½ Digits
6½ Digits
10 Ω
10.00000 Ω
1 mΩ
100 μΩ
10 μΩ
5 mA / 13 V
100 Ω
100.0000 Ω
10 mΩ
1 mΩ
100 μΩ
1 mA / 6 V
1 kΩ
1.000000 kΩ
100 mΩ
10 mΩ
1 mΩ
1 mA / 6 V
10 kΩ
10.00000 kΩ
1 Ω
100 mΩ
10 mΩ
100 μA / 6 V
100 kΩ
100.0000 kΩ
10 Ω
1 Ω
100 mΩ
10 μA / 13 V
1 MΩ
1.000000 MΩ
100 Ω
10 Ω
1 Ω
10 μA / 13 V
10 MΩ
10.00000 MΩ
1 kΩ
100 Ω
10 Ω
1 μA / 13 V
100 MΩ
100.0000 MΩ
10 kΩ
1 kΩ
100 Ω
1 μA || 10 MΩ / 10 V
1.0 GΩ
1.000000 GΩ
100 kΩ
10 kΩ
1 kΩ
1 μA || 10 MΩ / 10 V
DMM4050/4040 Accuracy
Range
24 Hour
(23 °C ±1 °C)
90 Days
(23 °C ±5 °C)
1 Year
(23 °C ±5 °C)
Temperature Coefficient/°C
Outside 18 to 28 °C
Accuracy is given as ±(% measurement + % of range)
10 Ω
0.003 + 0.01
0.008 + 0.03
0.01+ 0.03
0.0006 + 0.0005
100 Ω
0.003 + 0.003
0.008 + 0.004
0.01 + 0.004
0.0006 + 0.0005
1 kΩ
0.002 + 0.0005
0.008 + 0.001
0.01 + 0.001
0.0006 + 0.0001
10 kΩ
0.002 + 0.0005
0.008 + 0.001
0.01 + 0.001
0.0006 + 0.0001
100 kΩ
0.002 + 0.0005
0.008 + 0.001
0.01 + 0.001
0.0006 + 0.0001
1 MΩ
0.002 + 0.001
0.008 + 0.001
0.01 + 0.001
0.001 + 0.0002
10 MΩ
0.015 + 0.001
0.02 + 0.001
0.04 + 0.001
0.003 + 0.0004
100 MΩ
0.3 + 0.01
0.8 + 0.01
0.8 + 0.01
0.15 + 0.0002
1 GΩ
1.0 + 0.01
1.5 + 0.01
2.0 + 0.01
0.6 + 0.0002
Additional Ohms Errors
Digits
NPLC
Additional NPLC Noise Error
100
0% of range
10
0% of range
1
0.001% of range
0.2
0.003% of range ±7 mΩ
0.02
0.017% of range ±15 mΩ

DC CURRENT

Characteristic
Description
Input Protection
Tool-accessible 11 A / 1000 V and 440 mA / 1000 V fuses, limits of 400 mA continuous 550 mA for 2 minutes on, 1 minute off
Common Mode Rejection
140 dB at 50 or 60 Hz ±0.1% (1 kΩ unbalance)
Normal Mode Rejection
60 dB for NPLC of 1 or greater with analog filter off and power line frequency ±0.1%
100 dB for NPLC of 1 or greater with analog filter on and power line frequency ±0.1%
Analog Filter
When using the analog filter, specifications are relative to within one hour of using the ZERO function for that range and NPLC setting

Input Characteristics
Range
Resolution
Resolution
Shunt Resistance
(Ohms)
Burden Voltage
4½ Digits
5½ Digits
6½ Digits
100 μA
100.0000 μA
10 nA
1 nA
100 pA
100 Ω
<0.015 V
1 mA
1.000000 mA
100 nA
10 nA
1 nA
100 Ω
<0.15 V
10 mA
10.00000 mA
1 μA
100 nA
10 nA
1 Ω
<0.025 V
100 mA
100.0000 mA
10 μA
1 μA
100 nA
1 Ω
<0.25 V
400 mA 7
400.000 mA
100 μA
10 μA
1 μA
1 Ω
<0.50 V
1 A 6
1.000000 A
100 μA
10 μA
1 μA
0.01 Ω
<0.05 V
3 A 5
3.00000 A
1 mA
100 μA
10 μA
0.01 Ω
<0.15 V
10 A
10.00000 A
1 mA
100 μA
10 μA
0.01 Ω
<0.5 V
5 Part of 10 A range.
6 Available on the front-panel terminal only.
7 400 mA continuously; 550 mA for 2 minutes on, 1 minute off.
DMM4050/4040 Accuracy
Range
24 Hour
(23 °C ±1 °C)
90 Days
(23 °C ±5 °C)
1 Year
(23 °C ±5 °C)
Temperature Coefficient/°C
Outside 18 to 28 °C
Accuracy is given as ±(% measurement + % of range)
100 μA
0.01 + 0.02
0.04 + 0.025
0.05 + 0.025
0.002 + 0.003
1 mA
0.007 + 0.005
0.030 + 0.005
0.05 + 0.005
0.002 + 0.0005
10 mA
0.007 + 0.02
0.03 + 0.02
0.05 + 0.02
0.002 + 0.002
100 mA
0.01 + 0.004
0.03 + 0.005
0.05 + 0.005
0.002 + 0.0005
400 mA 7
0.03 + 0.004
0.04 + 0.005
0.05 + 0.005
0.005 + 0.0005
1 A 6
0.03 + 0.02
0.04 + 0.02
0.05 + 0.02
0.005 + 0.001
3 A 5, 6
0.05 + 0.02
0.08 + 0.02
0.1 + 0.02
0.005 + 0.002
10 A 6
0.1 + 0.008
0.12 + 0.008
0.15 + 0.008
0.005 + 0.0008
5 Part of 10 A range.
6 Available on the front-panel terminal only.
7 400 mA continuously; 550 mA for 2 minutes on, 1 minute off.
Additional Current Errors
Digits
NPLC
Additional NPLC Noise Error for 1 mA, 100 mA, 400 mA, 3 A, and 10 A
Additional NPLC Noise Error for 100 μA, 10 mA, 1 A
100
0% of range
0% of range
10
0% of range
0% of range
1
0.001% of range
0.01% of range
0.2
0.11% of range ±4 μA
0.11% of range ±4 μA
0.02
0.04% of range ±4 μA
0.28% of range ±4 μA

AC CURRENT

The following AC current specifications are for sinusoidal signals with amplitudes greater than 5% of range. For inputs from 1% to 5% of range, add an additional error of 0.1% of range.
Characteristic
Description
Input Protection
Tool-accessible 11 A / 1000 V and 440 mA / 1000 V fuses, limits of 400 mA continuous 550 mA for 2 minutes on, 1 minute off
Measurement Method
AC-coupled true RMS, DC-coupled to the fuse and shunt (no blocking capacitor)
AC Filter Bandwidth
   Slow
3 Hz to 10 kHz
   Medium
20 Hz to 10 kHz
   Fast
200 Hz to 10 kHz
Maximum Crest Factor
5:1 at Full Scale
Additional Crest Factor Errors (<100 Hz)
Crest factor 1-2, 0.05% of full scale
Crest factor 2-3, 0.2% of full scale
Crest factor 3-4, 0.4% of full scale
Crest factor 4-5, 0.5% of full scale
Only applies to non-sinusoid signals

Input Characteristics
Range
Resolution
Resolution
Shunt Resistance
(Ohms)
Burden Voltage
4½ Digits
5½ Digits
6½ Digits
100 μA
100.0000 μA
10 nA
1 nA
100 pA
100 Ω
<0.015 V
1 mA
1.000000 mA
100 nA
10 nA
1 nA
100 Ω
<0.15 V
10 mA
10.00000 mA
1 μA
100 nA
10 nA
1 Ω
<0.025 V
100 mA
100.0000 mA
10 μA
1 μA
100 nA
1 Ω
<0.25 V
400 mA 9
400.000 mA
100 μA
10 μA
1 μA
1 Ω
<0.50 V
1 A 8
1.000000 A
100 μA
10 μA
1 μA
0.01 Ω
<0.05 V
3 A 5, 8
3.00000 A
1 mA
100 μA
10 μA
0.01 Ω
<0.05 V
10 A 8
10.00000 A
1 mA
100 μA
10 μA
0.01 Ω
<0.5 V
5 Part of 10 A range.
8 Available at front-panel connectors only.
9 400 mA continuously; 550 mA for 2 minutes on, 1 minute off; maximum crest factor 3:1 at 400 mA.
DMM4050/4040 Accuracy
Range
Frequency
24 Hour
(23 °C ±1 °C)
90 Days
(23 °C ±5 °C)
1 Year
(23 °C ±5 °C)
Temperature Coefficient/°C
Outside 18 to 28 °C
Accuracy is given as ±(% measurement + % of range)
100 μA
3 – 5 Hz
1.1 + 0.06
1.1 + 0.06
1.1 + 0.06
0.2 + 0.006
5 – 10 Hz
0.35 + 0.06
0.35 + 0.06
0.35 + 0.06
0.1 + 0.006
10 Hz – 5 kHz
0.15 + 0.06
0.15 + 0.06
0.15 + 0.06
0.015 + 0.006
5 – 10 kHz
0.35 + 0.7
0.35 + 0.7
0.35 + 0.7
0.03 + 0.006
1 mA
3 – 5 Hz
1.0 + 0.04
1.0 + 0.04
1.0 + 0.04
0.1 + 0.006
5 – 10 Hz
0.3 + 0.04
0.3 + 0.04
0.3 + 0.04
0.035 + 0.006
10 Hz – 5 kHz
0.1 + 0.04
0.1 + 0.04
0.1 + 0.04
0.015 + 0.006
5 – 10 kHz
0.2 + 0.25
0.2 + 0.25
0.2 + 0.25
0.03 + 0.006
10 mA
3 – 5 Hz
1.1 + 0.06
1.1 + 0.06
1.1 + 0.06
0.2 + 0.006
5 – 10 Hz
0.35 + 0.06
0.35 + 0.06
0.35 + 0.06
0.1 + 0.006
10 Hz – 5 kHz
0.15 + 0.06
0.15 + 0.06
0.15+ 0.06
0.015 + 0.006
5 – 10 kHz
0.35 + 0.7
0.35 + 0.7
0.35 + 0.7
0.03 + 0.006
100 mA
3 – 5 Hz
1.0 + 0.04
1.0 + 0.04
1.0 + 0.04
0.1 + 0.006
5 – 10 Hz
0.3 + 0.04
0.3 + 0.04
0.3 + 0.04
0.035 + 0.006
10 Hz – 5 kHz
0.1 + 0.04
0.1 + 0.04
0.1 + 0.04
0.015 + 0.006
5 – 10 kHz
0.2 + 0.25
0.2 + 0.25
0.2 + 0.25
0.03 + 0.006
400 mA 7
3 – 5 Hz
1.0 + 0.1
1.0 + 0.1
1.0 + 0.1
0.1 + 0.006
5 – 10 Hz
0.3 + 0.1
0.3 + 0.1
0.3 + 0.1
0.035 + 0.006
10 Hz – 5 kHz
0.1 + 0.1
0.1 + 0.1
0.1 + 0.1
0.015 + 0.006
5 – 10 kHz
0.2 + 0.7
0.2 + 0.7
0.2 + 0.7
0.03 + 0.006
1 A 6
3 – 5 Hz
1.0 + 0.04
1.0 + 0.04
1.0 + 0.04
0.1 + 0.006
5 – 10 Hz
0.3 + 0.04
0.3 + 0.04
0.3 + 0.04
0.035 + 0.006
10 Hz – 5 kHz
0.1 + 0.04
0.1 + 0.04
0.1 + 0.04
0.015 + 0.006
5 – 10 kHz
0.35 + 0.7
0.35 + 0.7
0.35 + 0.7
0.03 + 0.006
3 A 5, 6
3 – 5 Hz
1.1 + 0.06
1.1 + 0.06
1.1 + 0.06
0.1 + 0.006
5 – 10 Hz
0.35 + 0.06
0.35 + 0.06
0.35 + 0.06
0.035 + 0.006
10 Hz – 5 kHz
0.15 + 0.06
0.15 + 0.06
0.15 + 0.06
0.015 + 0.006
5 – 10 kHz
0.35 + 0.7
0.35 + 0.7
0.35 + 0.7
0.03 + 0.006
10 A 6
3 – 5 Hz
1.1 + 0.06
1.1 + 0.06
1.1 + 0.06
0.1 + 0.006
5 – 10 Hz
0.35 + 0.06
0.35 + 0.06
0.35 + 0.06
0.035 + 0.006
10 Hz – 5 kHz
0.15 + 0.06
0.15 + 0.06
0.15 + 0.06
0.015 + 0.006
5 – 10 kHz
0.35 + 0.7
0.35 + 0.7
0.35 + 0.7
0.03 + 0.006
5 Part of 10 A range.
6 Available on the front-panel terminal only.
7 400 mA continuously; 550 mA for 2 minutes on, 1 minute off.
Additional Low-frequency Errors
Frequency
AC Filter
3 Hz (Slow)
20 Hz (Medium)
200 Hz (Fast)
Error is stated as % of reading.
10 – 20 Hz
0
0.25
20 – 40 Hz
0
0.02
40 – 100 Hz
0
0.01
0.55
100 – 200 Hz
0
0
0.2
200 Hz – 1 kHz
0
0
0.02
>1 kHz
0
0
0

FREQUENCY

Characteristic
Description
Gate Time
Programmable to 1 s, 100 ms, and 10 ms
Measurement Method
Flexible counting technique. AC-coupled input using the AC voltage measurement function
Settling Considerations
When measuring frequency or period after a DC offset voltage change, errors may occur. For the most accurate measurement, wait up to 1 second for the input blocking capacitor to settle
Measurement Considerations
To minimize measurement errors, shield inputs from external noise when measuring low-voltage, low-frequency signals
DMM4050/4040 Accuracy
Range
Frequency
24 Hour
(23 °C ±1 °C)
90 Days
(23 °C ±5 °C)
1 Year
(23 °C ±5 °C)
Temperature Coefficient/°C
Outside 18 to 28 °C
Accuracy is given as ± % measurement
100 mV to 1000 V10, 11
3 – 5 Hz
0.1
0.1
0.1
0.005
5 – 10 Hz
0.05
0.05
0.05
0.005
10 – 40 Hz
0.03
0.03
0.03
0.001
40 Hz – 300 kHz
0.006
0.01
0.01
0.001
300 kHz – 1 MHz
0.006
0.01
0.01
0.001
10 Limited to 8 × 107 Volt-Hertz
11 Input >100 mV. For 10 – 100 mV, multiply percent measurement error by 10.
Gate Time vs. Resolution
Gate Time
Resolution
0.01
0.1
1.0
Additional Low-frequency Errors
Frequency
NPLC
Error stated as percent of measurement for inputs >100 mV. For 10 – 100 mV, multiply percent by 10.
3 – 5 Hz
0
0.12
0.12
5 – 10 Hz
0
0.17
0.17
10 – 40 Hz
0
0.2
0.2
40 – 100 Hz
0
0.06
0.21
100 – 300 Hz
0
0.03
0.21
300 Hz – 1 kHz
0
0.01
0.07
>1 kHz
0
0
0.02
Capacitance (DMM4050 Only)
Range
Resolution
1 Year Accuracy 12
(23 °C ±5 °C)
Temperature Coefficient/°C
Outside 18 to 28 °C
Accuracy is stated as ±(% of measurement + % of range)
1 nF
1 pF
2% ±2.5%
0.05 + 0.05
10 nF
10 pF
1% ±0.5%
0.05 + 0.01
100 nF
100 pF
1% ±0.5%
0.01 + 0.01
1 μF
1 nF
1% ±0.5%
0.01 + 0.01
10 μF
10 nF
1% ±0.5%
0.01 + 0.01
100 μF
100 nF
1% ±0.5%
0.01 + 0.01
1 mF
1 μF
1% ±0.5%
0.01 + 0.01
10 mF
10 μF
1% ±0.5%
0.01 + 0.01
100 mF
100 μF
4% ±0.2%
0.05 + 0.05
12 Stated accuracy is attained when Zero function is used.
Temperature (DMM4050 only)
Range
Resolution
Accuracy
Temperature Coefficient/°C
Outside 18 to 28 °C
90 Days
(23 °C ±5 °C)
1 Year
(23 °C ±5 °C)
Test Current: 1 mA
Accuracy is stated as ± °C and is based on a Platinum RT100 (DIN IEC 751, 385 type) RTD with less than 10 Ω lead resistance. The accuracy listed in the table below are valid only when using the 4-wire RTD measurement function. Specifications do not include probe accuracy, which must be added.
–200 °C
0.001 °C
0.06
0.09
0.0025
–100 °C
0.001 °C
0.05
0.08
0.002
0 °C
0.001 °C
0.04
0.06
0.002
100 °C
0.001 °C
0.05
0.08
0.002
300 °C
0.001 °C
0.1
0.12
0.002
600 °C
0.001 °C
0.18
0.22
0.002
Additional Errors
Digits
NPLC
Additional NPLC Noise Error
100
0 °C
10
0 °C
1
0.03 °C
0.2
0.12 °C
0.02
0.6 °C

CONTINUITY

Characteristic
Description
Continuity Threshold
Selectable between 1 Ω and 1000 Ω
Test Current
1 mA
Response Time
300 S/s with audible tone
Range
24 Hour
(23 °C ±1 °C)
90 Days
(23 °C ±5 °C)
1 Year
(23 °C ±5 °C)
Temperature Coefficient/°C
Outside 18 to 28 °C
Accuracy is given as ±(% measurements + % of range)
1000.0 Ω
0.002 + 0.01
0.008 + 0.02
0.01 + 0.02
0.001 + 0.002

DIODE TEST

Characteristic
Description
Test Current
100 μA or 1 mA
Response Time
300 S/s with audible tone
Range
24 Hour
(23 °C ±1 °C)
90 Days
(23 °C ±5 °C)
1 Year
(23 °C ±5 °C)
Temperature Coefficient/°C
Outside 18 to 28 °C
Accuracy is given as ±(% measurements + % of range)
5.0000 V
0.002 + 0.002
0.008 + 0.002
0.01 + 0.002
0.001 + 0.002
10.0000 V
0.002 + 0.001
0.008 + 0.002
0.01 + 0.002
0.001 + 0.002
Measurement Rates (IEEE48816)
Function
Digits
Setting
Integration Time
60 Hz (50 Hz)
Measurements/Second13
DMM4040
DMM4050
DC Volts, DC Current, and Resistance
100 NPLC
1.67 (2) s
0.6 (0.5)
0.6 (0.5)
10 NPLC
167 (200) ms
6 (5)
6 (5)
1 NPLC
16.7 (20) ms
60 (50)
60 (50)
0.2 NPLC
3.3 ms
270
270
0.02 NPLC
500 μs
995
995
AC Voltage and AC Current14
3 Hz

0.47
0.47
20 Hz

1.64
1.64
200 Hz15

4.5
4.5
Frequency and Period
1 s

1
1
100 ms

9.8
9.8
10 ms

80
80
Capacitance


NA
2
13 Typical measurement rates with auto-zero off, delay = 0, display off, auto range off and math off.
14 Maximum measurement rates for 0.01% of AC step. When DC input varies, additional settling delay is required.
15 For remote operation or external trigger using default settling delay.
16 Note that the measurement rates for RS232 can vary depending on the baud rate chosen. If the baud rate selected is 115,200, the maximum measurement rate is 711 measurements. The LAN bus has a maximum measurement rate of 963 measurements

Không có nhận xét nào: